163
Chapter 4 Remote Interface Reference
Using the APPLy Command
4
Output Frequency
•For the frequency parameter of the APPLy command, the output
frequency range depends on the function specified. You can substitute
“MINimum”, “MAXimum”, or “DEFault” in place of a specific value
for the frequency parameter. MIN selects the lowest frequency
allowed for the function specified and MAX selects the highest
frequency allowed. The default frequency is 1 kHz for all functions.
Limits Due to Function: The frequency limits are determined by the
function specified in the APPLy command. The APPLy command
always sets both the function and the frequency, and the specified
frequency must be appropriate for the function. For example, the
command APPL:RAMP 10 MHz results in a Data out of range” error.
The frequency is set to 100 kHz, which is the maximum for a ramp
waveform.
Function Minimum Frequency Maximum Frequency
Sine
Square
Ramp
Pulse
Noise, DC
Arbs
1 mHz
1 mHz
1 mHz
1 mHz
Not Applicable
1 mHz
10 MHz
10 MHz
100 kHz
5 MHz
Not Applicable
3 MHz
33210A users guide.book Page 163 Wednesday, July 16, 2008 11:16 AM