46
Chapter 2  Front-Panel Menu Operation
To Configure the Remote Interface
2
To Configure the Remote Interface 
The Keysight 33210A supports remote interface communication using a
choice of three interfaces: GPIB, USB, and LAN (LXI Class C compliant).
All three interfaces are "live" at power up. The following sections tell how
to configure the remote interface from the instrument front panel.
Note: Two CDs, provided with your instrument, contain connectivity 
software to enable communications over the remote interfaces. See 
“Connectivity Software and Product CDs,” page 128, for further 
information on these CDs and the software they contain. 
GPIB Configuration 
You need only select a GPIB address.
1 Select the “I/O” menu.
Press and then press the 
I/O
 
softkey. 
2 Set the GPIB address.
Use the knob and cursor keys or the numeric keypad to select a GPIB
address in the range 0 through 30 (the factory default is “10”). 
The GPIB address is shown on the front-panel display at power-on.
3 Exit the menu.
Press the DONE softkey.
USB Configuration
The USB interface requires no front panel configuration parameters.
Just connect the 
Keysight 33210A to your PC with the appropriate USB
cable. The interface will self configure. Press the Show USB Id softkey in
the “I/O menu” to see the USB interface identification string. Both USB 
1.1 and USB 2.0 are supported.
33210A users guide.book Page 46 Wednesday, July 16, 2008 11:16 AM