237
Chapter 4  Remote Interface Reference
System-Related Commands
4
System-Related Commands 
See also “System-Related Operations” starting on page 119 in chapter 3. 
SYSTem:ERRor? 
Read and clear one error from the function generator’s error queue. 
A record of up to 20 command syntax or hardware errors can be stored in 
the error queue. See chapter 5 for a complete listing of the error messages. 
• Errors are retrieved in first-in-first-out (FIFO) order. The first error 
returned is the first error that was stored. Errors are cleared as you 
read them. The function generator beeps once each time an error is 
generated (unless disabled using the 
SYST:BEEP:STAT
 command)
.
• If more than 20 errors have occurred, the last error stored in the 
queue (the most recent error) is replaced with “Queue overflow”. 
No additional errors are stored until you remove errors from the 
queue. If no errors have occurred when you read the error queue, 
the function generator responds with “No error”. 
• The error queue is cleared by the *CLS (clear status) command or 
when the power is cycled. The errors are also cleared when you read 
the queue. The error queue is not cleared by a reset (*RST command). 
• Errors have the following format (the error string may contain up to 
255 characters). 
-113,"Undefined header" 
33210A users guide.book  Page 237  Wednesday, July 16, 2008  11:16 AM