Chapter 5 Error Messages
Execution Errors
5
Execution Errors
-211 Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was
ignored. Make sure that you have selected the proper trigger source and
verify that the sweep or burst mode is enabled.
-221 Settings conflict;
Option 002 (arbs) is not installed
Arbitrary waveforms are optionally available with the 33210A (Option
002). To upgrade your 33210A to include arbitrary waveform
functionality, go to www.
keysight.com/find/33210U.
-221 Settings conflict;
turned off infinite burst to allow immediate trigger source
An infinite count burst is allowed only when the external or bus
(software) trigger source is selected. The burst count has been set to the
maximum N Cycle value (50,000 cycles).
-221 Settings conflict;
infinite burst changed trigger source to BUS
An infinite count burst is allowed only when the external or bus
(software) trigger source is selected. Sending the BURS:NCYC INF
command has automatically changed the trigger source from immediate
to bus.
-221 Settings conflict;
burst period increased to fit entire burst
The number of cycles specified by the BURS:NCYC command takes
priority over the burst period (as long as the burst period is not at its
maximum value). The function generator has increased the burst period
to accommodate the specified burst count or waveform frequency.
-221 Settings conflict;
burst count reduced to fit entire burst
Since the burst period is currently at its maximum, the function generator
has reduced the burst count to allow the specified waveform frequency.