259
Chapter 4  Remote Interface Reference
Status Reporting Commands
4
Status Reporting Commands 
Use the following commands to access the Status System Registers.
See the table on page 251 for the register bit definitions.
*STB? 
Query the summary (condition) register in this register group. 
This command is similar to a Serial Poll but it is processed like any other 
instrument command. This command returns the same result as a 
Serial Poll but the “Master Summary” bit (bit 6) is not cleared by the 
*STB? command.
*SRE <enable value> 
*SRE? 
Enable bits in the Status Byte to generate a Service Request. To enable 
specific bits, you must write a decimal value which corresponds to the 
binary-weighted sum of the bits in the register. The selected bits are 
summarized in the “Master Summary” bit (bit 6) of the Status Byte 
Register. If any of the selected bits change from “0” to “1”, a Service 
Request signal is generated. The *SRE? query returns a decimal value 
which corresponds to the binary-weighted sum of all bits enabled by the 
*SRE command.
•A *CLS (clear status) does not clear the enable register but it does 
clear all bits in the event register. 
•A STATus:PRESet does not clear the bits in the Status Byte 
enable register.
•A *PSC 0 preserves the contents of the enable register through 
power cycles. 
Status Byte Register Commands
33210A users guide.book  Page 259  Wednesday, July 16, 2008  11:16 AM