67
Chapter 4 Calibration Procedures
+10 dB Range Flatness Verification
4
4
+10 dB Range Flatness Verification
This procedure checks the high frequency ac amplitude flatness above
100 kHz on the +10dB attenuator range. (Flatness is relative to 1 kHz.)
1 Connect the power meter to measure the output amplitude of the
instrument as shown on page 65.
2 Set up the function generator as follows:
Output impedance: 50 (press and select Output Setup).
•Waveform: Sine
Frequency: 100 kHz
Amplitude: 13.00 dBm
Make sure the output is enabled.
3 On the power meter, use the Relative Power function to set the current
reading as the reference value. This will allow you to compare future
measurement results in dB.
4 Set the power meter offset to equal the 100kHz_10dB_offset value
previously calculated. This sets the power meter to directly read the
flatness error specification relative to 1 kHz. 100kHz_10dB_offset is
calculated on the Amplitude and Flatness Verification Worksheet.