92
Chapter 4 Calibration Procedures
Calibration Errors
4
Calibration Errors
The following errors are failures that may occur during a calibration. 
System error messages are described in the Keysight 33210A User’s 
Guide. Self-test error messages are described beginning on page 110 of 
this document
701 Calibration error; security defeated by hardware jumper
The function generator’s calibration security feature has been disabled 
by temporarily shorting the two “CAL ENABLE” pads on the internal 
circuit board as described starting on page 73.
702 Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured. 
See “To Unsecure and Secure for Calibration”, on page 41 for a procedure 
to enter the security code from the front panel. Use the CAL:SEC:STAT 
ON command to enter the security code using the remote interface.
703 Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command.
706 Calibration error; value out of range
You have entered a value that was unexpected by the calibration firmware. 
For example, if a number is expected such a 50.XX ohms, and you enter 
10 ohms, that number is outside the expected range of valid inputs.
707 Calibration error; signal input is out of range
Occurs during the ADC Adjustment, setup 6, if the 1 Volt input voltage is 
too high. May also occur during self-calibration (setup 7), run self-test to 
diagnose cause of problem.
707 707: Calibration error; cal edge time; rise time cal error
707: Calibration error; cal edge time; fall time cal error
707: Calibration error; cal edge time; default values loaded
Indicates a failure in the rise-time or fall-time circuitry has prevented 
calibration. The edge-time was calibrated using default values, limiting 
accuracy. Service is required to correct the problem and achieve design 
accuracy for the rise and fall times.