87
Chapter 4  Calibration Procedures
+10 dB Range Flatness Adjustments
4
4
+10 dB Range Flatness Adjustments
Note The Linear Phase path is not adjusted. It is approximated using the 
other path’s values.
1 Connect the power meter as shown below. 
2 Use a power meter to measure the output amplitude for each of the 
setups in the table on the next page. 
Note Setup 54 establishes the power meter reference for all the remaining 
setups in this table. You must always perform setup 54 before any of the 
following setups.