Chapter 4  Calibration Procedures
AC Amplitude (high-impedance) Adjustment
4
4
AC Amplitude (high-impedance) Adjustment
The function generator stores a calibration constant for each high-
impedance attenuator path. The gain coefficient of each path is 
calculated using two measurements; one with the waveform DAC at + 
output and one with waveform DAC at – output. The setups, therefore, 
must be performed in pairs. 
1 Connect the DMM as shown below.