56
Chapter 4 Calibration Procedures
Recommended Test Equipment
4
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not
available, substitute calibration standards of equivalent accuracy.
* Q = Quick Verification P = Performance Verification T = Troubleshooting
** An oscilloscope is not required for calibration, only for troubleshooting.
Instrument Requirements Recommended Model Use*
Digital Multimeter
(DMM)
ac volts, true rms, ac coupled
accuracy: ±0.02% to 1 MHz
dc volts
accuracy: 50 ppm
resolution: 100 µV
Resistance
Offset-compensated
accuracy: ±0.1Ω
Keysight 3458A Q, P, T
Power Meter 100 kHz to 100 MHz
1 µW to 100 mW (–30 dBm to +20 dBm)
accuracy: 0.02 dB
resolution: 0.01 dB
Keysight E4418B
Q, P, T
Power Head 100 kHz to 100 MHz
1 µW to 100 mW (–30 dBm to +20 dBm)
Keysight 8482A
Q, P, T
Attenuator –20 dB
Keysight 8491A Opt 020
Q, P, T
Frequency Meter accuracy: 0.1 ppm
Keysight 53131A Opt
010 (high stability)
Q, P, T
Oscilloscope** 500 MHz
2 Gs/second
50Ω input termination
Keysight 54831B
T
Adapter N type (m) to BNC (m) N type (m) to BNC (m) Q, P, T
Cable BNC (m) to dual-banana (f)
Keysight 10110B
Q, P, T
Cable (2 required) Dual banana (m) to dual banana (m)
Keysight 11000-60001
Q, P, T
Cable RG58, BNC (m) to dual banana
Keysight 11001-60001
Q, P, T
Cable RG58, BNC (m) to BNC (m)
Keysight 8120-1840
Q, P, T