Chapter 4 Calibration Procedures
0 dB Range Flatness Adjustments
4
4
0 dB Range Flatness Adjustments
1 Connect the power meter as shown on page 87.
2 Use the power meter to measure the output amplitude for each of the
setups in the table below.
Note Setup 39 establishes the power meter reference for all the remaining
setups in this table. You must always perform setup 39 before any of the
following setups.
* Constants are stored after completing this setup.
Nominal Signal
Setup Frequency Amplitude
39* 100 kHz 0.28 Vrms 2 dBm Power Meter Reference for 0dB Range
40* 200 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
41* 500 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
42* 1.5 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
43* 3 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
44* 4 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
45* 6 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
46* 8 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
47* 9.9 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
48* 200 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
49* 500 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
50* 1.5 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
51* 3.0 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
52 Setup not used for this instrument
53 Setup not used for this instrument