80
Chapter 4 Calibration Procedures
Output Impedance Adjustment
4
2 Use the DMM to make a resistance measurement at the front panel
Output connector for each setup in the following table. The expected
measured value is approximately 50 Ω.
* Constants are stored after completing this setup.
3 Using the numeric keypad or knob, adjust the displayed impedance at
each setup to match the measured impedance. Select ENTER VALUE.
4 There are no specific operational verification tests for Output
Impedance. Continue with the next adjustment procedure in this
chapter.
Setup
8* -30dB range
9* -20dB range
10* -10dB range
11* 0dB range
12* +10dB range