Chapter 4 Calibration Procedures
0 dB Range Flatness Verification
4
5 Set the function generator to each output described in the table below
and measure the output amplitude with the power meter (the relative
measurement in dB).
6 Compare the measured output to the test limits shown in the table.
Keysight 33210A
Measurement
Output Setup Function Amplitude Frequency Nominal Error
Q 50 Ω Sine Wave +3.51 dBm 100.000 kHz 0 dB ±0.1 dB
50 Ω Sine Wave +3.51 dBm 200.000 kHz 0 dB ±0.2 dB
50 Ω Sine Wave +3.51 dBm 500.000 kHz 0 dB ±0.2 dB
50 Ω Sine Wave +3.51 dBm 2.000 MHz 0 dB ±0.2 dB
50 Ω Sine Wave +3.51 dBm 3.000 MHz 0 dB ±0.2 dB
Q 50 Ω Sine Wave +3.51 dBm 4.000 MHz 0 dB ±0.2 dB
50 Ω Sine Wave +3.51 dBm 5.000 MHz 0 dB ±0.2 dB
50 Ω Sine Wave +3.51 dBm 8.000 MHz 0 dB ±0.3 dB
Q 50 Ω Sine Wave +3.51 dBm 10.000 MHz 0 dB ±0.3 dB