3 Adjustment Procedures
50 Keysight 3458A Calibration Manual
11 Set the synthesizer to output 3 V rms at 100 kHz. Execute the
SCAL 1E5 command. The multimeter automatically performs the adjustment.
When the adjustment is complete, the multimeter returns to displaying DC
voltage readings.
12 Without changing the synthesizer settings, execute the SCAL 10 command as
shown on Line B of the Adjustment Record.
13 Set the synthesizer to the voltage and frequency shown on Line C of the
Adjustment Record. Execute the SCAL command as shown on Line C of the
Adjustment Record.
14 Repeat step 13 for each synthesizer setting and SCAL command shown on
Lines D through H on the Adjustment Record.
15 Disconnect all equipment from the multimeter.
16 Execute the ACAL AC command.