206 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
11 Triggers
 
When two non-overlapping zones are used, their conditions are ANDed to become 
the final qualifying condition.
When two overlapping zones have the must intersect condition, the zones are 
ORed. When two overlapping zones have different conditions, Zone 1 takes 
precedence and Zone 2 is not used. In this case, Zone 2 will have no fill (that is, 
neither solid nor shaded) to indicate that it is not being used.
The zone qualified trigger feature is incompatible with, and will disable, the XY and 
Roll horizontal time modes and the Averaging acquisition mode.
Keep in mind that the TRIG OUT signal comes from the oscilloscope's hardware trigger. The 
TRIG OUT signal indicates when there is a trigger (acquisition) that is evaluated for zone 
intersection, not when an acquisition meets the zone qualification and is plotted on the 
oscilloscope's display.