Keysight InfiniiVision
DSOX1204A/G
Oscilloscopes
Service Guide

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Keysight DSOX1204A and is the answer not in the manual?

Keysight DSOX1204A Specifications

General IconGeneral
BrandKeysight
ModelDSOX1204A
CategoryTest Equipment
LanguageEnglish

Summary

Testing Performance

To verify DC vertical gain accuracy

Details the procedure for testing the DC vertical gain accuracy for each analog channel.

To verify bandwidth (-3 dB)

Explains the test procedure for verifying the oscilloscope's bandwidth using a signal generator.

To verify time base accuracy

Describes the test procedure for verifying the accuracy of the oscilloscope's time base oscillator.

To verify trigger sensitivity

Details tests for verifying the oscilloscope's trigger sensitivity at different bandwidths.

Calibrating and Adjusting

User Calibration

Explains the conditions and process for performing user calibration for optimal performance.

To perform User Cal

Provides step-by-step instructions for initiating the user calibration routine.

Troubleshooting

Solving General Problems with the Oscilloscope

Guides users through common issues like no display, no trace, or unexpected display.

Verifying Basic Operation

Outlines procedures to check the oscilloscope's basic functionality, including power-on and self-tests.

Safety Notices

Warnings

Provides critical safety warnings regarding power connection, fuses, instrument modification, and environment.