EasyManua.ls Logo

Keysight DSOX1204A - Maximum Input Voltages; Environmental Conditions

Keysight DSOX1204A
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Characteristics and Specifications 1
DSOX1204A/G Oscilloscopes Service Guide 13
Maximum Input Voltages
Environmental Conditions
CAUTION
Maximum input voltage for analog inputs:
150 Vrms, 200 Vpk
CAUTION
Maximum input voltage for external trigger input:
30 Vrms
Environment Indoor use only.
Ambient
temperature
Operating: 0
°C to +50 °C
Non-operating: –40 °C to +70 °C
Humidity Operating: Up to 95% RH, non-condensing to temperatures up to +40 °C
decreasing linearly to 50% RH at +50 °C
Non-operating: Up to 90% RH up to +65 °C (non condensing)
Altitude Operating: Up to 3,000 m
Non-operating: Up to 15,300 m
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision DSOX1204A/G oscilloscopes may be operated in environments
of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.