Electrostatic Discharge (ESD) precautions
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component
damage can occur at electrostatic discharge voltages as low as 50 V.
The following guidelines will help prevent ESD damage during service operations:
Disassemble instruments only in a static-free work area.
Use a conductive work area to reduce static charges.
Use a conductive wrist strap to reduce static charge accumulation.
Minimize handling.
Keep replacement parts in original static-free packaging.
Remove all plastic, foam, vinyl, paper, and other static-generating materials from the immediate work area.
Troubleshooting
Before troubleshooting or repairing the instrument, make sure the failure is in the instrument rather than any
external connections. Also make sure that the instrument was accurately calibrated within the last year (see
Calibration Adjustment Procedures > Calibration Interval for details).
Perform the following verifications if the unit is inoperative:
Verify that the ac power cord is connected to the power supply.
Verify that the front-panel power switch is depressed.
Verify that the correct power-line fuse is installed. See To Replace the Power Line Fuse for details.
Verify the power-line voltage setting.
Self-Test Procedures
A power-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the
power supply is operational.
Press Utilities > Test / Setup > Self Test to perform the complete self-test of the power supply. It takes
approximate 2 seconds for the self-test to complete.
You can also perform a complete self-test from the remote interface, see Programming Guide for details.
If the self-test is successful, "Self test passed" is displayed on the front panel.
If the self-test fails, "!Err" is displayed on the front panel. Record the error code and message and contact Keysight
support if necessary.
If the self-test is successful, this indicates a high chance that the power supply is operational.
Keysight E36300 Series Service Guide 13

Table of Contents

Other manuals for Keysight E36300 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E36300 Series and is the answer not in the manual?

Keysight E36300 Series Specifications

General IconGeneral
BrandKeysight
ModelE36300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals