Test record form - Keysight E36313A
E36313A Report Number ____________ Date ______________
Description Outputs Lower limit Result Upper
limit
Constant Voltage Tests
Voltage Programming
Zero Voltage Output (V
0
)
Maximum Voltage Ouput (V
max
)
Output 1
Output 2
Output 3
Output 1
Output 2
Output 3
-0.003 V
-0.005 V
-0.005 V
5.9952 V
24.9875 V
24.9875 V
______________
______________
______________
______________
______________
______________
0.003 V
0.005 V
0.005 V
6.0048
V
25.0125
V
25.0125
V
Voltage Readback
Zero Voltage measured over interface
Maximum Voltage measured over interface
Output 1
Output 2
Output 3
Output 1
Output 2
Output 3
V
0
- 0.003 V
V
0
- 0.005 V
V
0
- 0.005 V
V
max
- 0.0054 V
V
max
- 0.0125 V
V
max
- 0.0125 V
______________
______________
______________
______________
______________
______________
V
0
+
0.003 V
V
0
+
0.005 V
V
0
+
0.005 V
V
max
+
0.0054
V
V
max
+
0.0125
V
V
max
+
0.0125
V
CV Load Regulation (V
load
- V
noload
) Output 1
Output 2
Output 3
-4.6 mV
-6.5 mV
-6.5 mV
______________
______________
______________
4.6 mV
6.5 mV
6.5 mV
CV Line Regulation (V
lowline
- V
highline
) Output 1
Output 2
Output 3
-1.6 mV
-3.5 mV
-3.5 mV
______________
______________
______________
1.6 mV
3.5 mV
3.5 mV
40
Keysight E36300 Series Service Guide