Keysight E4991B Impedance Analyzer 17
Installation
Power Supply
1-
Power Supply
Before turning on the E4991B power, check the following.
Verification of the Power Supply
Confirm that the power supplied to the E4991B meets the following
requirements:
Verification and Connection of Power Cable
The three-wire power cable attached to the E4991B has one wire serving as
the ground. Using this power cable allows the E4991B to be grounded, thereby
protecting you against any potential electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 1-5 on page 22) on the rear panel of the E4991B and a three-wire
power outlet with the grounding prong firmly connected in the ground slot.
Power cord list, 16000-99101 shows the power cable options.
Requirements
Rated voltage 100 - 240 VAC
Voltage range 90 - 264 VAC
Rated frequency 50 / 60 Hz
Frequency range 47 - 63 Hz
Maximum power consumption 300 VA
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Use the supplied power cable with grounding wire to securely ground the
E4991B.