Troubleshooting 1
Impedance Analyzer 13
How to exit from the E4991B Measurement View
Some troubleshooting procedures require you to exit from the
Keysight E4991B Measurement View. The following procedure
describes the steps to to exit from the Keysight E4991B
Measurement View.
Step 1. Connect a mouse to the connectors on the Keysight
E4991B rear panel.
Step 2. Turn the instrument on.
Step 3. Press key.
Step 4. Click Service.
Step 5. Click Exit.
Step 6. Click Yes in Exit menu. The Keysight E4991B exits from
Measurement View to Windows desktop.
Figure 1 Exit dialog box
NOTE
If you want to return to the Measurement View, double-click the
Keysight E4991B” icon on the desktop.
NOTE
if you need to shut down the Keysight E4991B, press the standby
switch on the front panel.