28 Keysight Infiniium S-Series Oscilloscopes User’s Guide
2 Using the Oscilloscope
AUX OUT
This output signal is selected by the Infiniium oscilloscope application's
Calibration Output dialog box. It can be a DC level, the probe compensation signal
(a square wave used to adjust compensated passive probes), the trigger out
signal, or a demo signal.
10 MHz REF IN 50
The 10 MHz REF IN 50 BNC connector is used to synchronize the oscilloscope's
horizontal timebase system to a reference clock that you provide.
The clock you provide must meet the following specifications:
Input frequency lock range: 10 MHz ±20 ppm
Amplitude, sine wave input: 356 mVpp (-5 dBm) min to 5 Vpp (+18 dBm) max
Amplitude, square wave input: 285 mVpp min to 4 Vpp max
Input impedance: 50 (typical)
To use an external reference clock, connect the external clock to the
10 MHz REF IN 50 BNC connector; then, in the Infiniium oscilloscope
application's Horizontal dialog box (Setup > Horizontal...), enable the External
10 MHz Reference Clock.
10 MHz REF OUT 50
You can use the 10 MHz REF OUT 50 BNC connector to send the oscilloscope’s
10 MHz reference clock output signal to another instrument’s reference clock
input.
The output has these characteristics:
Amplitude into 50 (internal or external timebase reference selected): 1.65 ±0.05 Vpp
(8.3 ±0.3 dBm) sine wave
Frequency accuracy, internal timebase reference selected: 10 MHz ±(12 ppb initial +
75 ppb/year aging)
Frequency accuracy, external timebase reference selected: external reference
frequency (input to 10 MHz REF IN 50)
AUX TRIG IN 50
You can set up the oscilloscope to trigger on the auxiliary trigger signal connected
to this BNC input.
The input impedance is 50, and the signal you provide must be 5 Vpp maximum
between -5 V and +5 V.

Table of Contents

Other manuals for Keysight Infiniium MSOS804A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Infiniium MSOS804A and is the answer not in the manual?

Keysight Infiniium MSOS804A Specifications

General IconGeneral
BrandKeysight
ModelInfiniium MSOS804A
CategoryTest Equipment
LanguageEnglish

Related product manuals