Testing Performance 6
Keysight Infiniium MXR/EXR-Series Real-Time Oscilloscopes User's Guide 79
Analog Bandwidth—Maximum Frequency Test
Specifications
Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
The oscilloscope must be warmed up (with the oscilloscope application running) for at least
30 minutes at ambient temperature prior to the start of any performance test.
Analog Bandwidth (-3 dB)
MXR608A, MXR604A 6 GHz
MXR408A, MXR404A 4 GHz
MXR258A, MXR254A, EXR258A, EXR254A 2.5 GHz
MXR208A, MXR204A, EXR208A, EXR204A 2 GHz
MXR108A, MXR104A, EXR108A, EXR104A 1 GHz
MXR058A, MXR054A, EXR058A, EXR054A 500 MHz