EasyManua.ls Logo

Keysight Infiniium MXR Series - Page 91

Keysight Infiniium MXR Series
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing Performance 6
Keysight Infiniium MXR/EXR-Series Real-Time Oscilloscopes User's Guide 91
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 2
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 3
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 4
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Table 8 Offset Accuracy Test (50 ) (continued)
Vertical Scale Ext.
Supply
Setting
V
DMM
+
V
Scop
e+
V
error
+
V
DMM
-
V
Scop
e-
V
error-
Offset
Accuracy
Limit (±)
V
zero-error
V
zero-error
Limit (±)

Table of Contents

Related product manuals