96 Keysight Infiniium MXR/EXR-Series Real-Time Oscilloscopes User's Guide
6 Testing Performance
5V/div ±40V 1.102V 502mV
Channel 8
5 mV/div ±2 V 32.5 mV 2.5 mV
10 mV/div ±5 V 78.0 mV 3.0 mV
20 mV/div ±10 V 154.0 mV 4.0 mV
50 mV/div ±10 V 157.0 mV 7.0 mV
100 mV/div ±20 V 312.0 mV 12.0 mV
200 mV/div ±20 V 322.0 mV 22.0 mV
500 mV/div ±20 V 352.0 mV 52.0 mV
1V/div ±40V 702.0mV 102.0 mV
2V/div ±40V 802.0mV 202mV
5V/div ±40V 1.102V 502mV
Table 9 Offset Accuracy Test (1 M) (continued)
Vertical Scale Ext.
Supply
Setting
V
DMM
+
V
Scop
e+
V
error
+
V
DMM
-
V
Scop
e-
V
error-
Offset
Accuracy
Limit (±)
V
zero-error
V
zero-error
Limit (±)
Table 10 DC Gain Accuracy Test (50 )
Vertical
Scale
Ext. Supply
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calculated
DC Gain
Error
DC Gain
Error Test
Limits
Channel 1
5 mV/div ±15 mV ±2 %
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1V/div ±3.0V ±2 %
Channel 2
5 mV/div ±15 mV ±2 %
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %