EasyManua.ls Logo

Keysight N6705 - Page 72

Keysight N6705
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Calibration
72 Model N6705 Service Guide
Front Panel:
SCPI Command:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
CAL:DATA <data>
10 μA Measurement Range - Models N6781A, N6782A, N6784A
Step 1.
Connect the Keysight 3458A in series with a 1 MΩ resistor to the output
(see equipment list in chapter 2 and Cal Setup D). Set the Keysight
3458A to measure current using the 100 μA range (no autoranging). The
remote command to do this is “DCI 100e-6”.
Step 2.
Select the 10 μA measurement range of output 1.
Front Panel:
SCPI Command:
From the Calibration menu, select
Current, 10 μA Measurement
Range. Then select Start.
CAL:CURR:MEAS 0.00001, (@1)
Step 3.
Select the first current calibration point.
Front Panel:
SCPI Command:
The first calibration point is automatically
selected. “Enter P1 measured data” should be
displayed.
CAL:LEV P1
*OPC?
Step 4.
Measure the current with the
Keysight 3458A and enter the value.
Front Panel:
SCPI Command:
Enter the value from the external DMM into
the data field and then select Next.
CAL:DATA <data>
Step 5.
Select the second current calibration point.
Front Panel:
SCPI Command:
The second calibration point is automatically
selected. “Enter P2 measured data” should be
displayed.
CAL:LEV P2
*OPC?
Step 6.
Measure the current with the Keysight 3458A and enter the value. This
value should be negative.
Front Panel:
SCPI Command:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
CAL:DATA <data>

Related product manuals