Calibration 3
Model N6705 Service Guide 73
Current Limit All Models
CAUTION
This procedure calibrates the POSITVE current limit for Models N6781A,
N6782A, and N6784A. It calibrates the NEGATIVE current limit for
Models N6785A and N6786A.
Step 1.
Connect a precision shunt resistor to the output. The shunt resistor
should be able to measure at least 120% of the output’s rated full-scale
current (see Cal Setup B). Connect the Keysight 3458A across the shunt
resistor.
Step 2.
Select the 3 A current limit of output 1 for models N6781A, N6782A, and
N6784A. Select the -8A current limit of output 1 for models N6785A and
N6786A.
Front Panel:
SCPI Command:
From the Calibration menu, select
Current, Current Limit. Then select
Start.
For N6781A, N6782A, N6784A
CAL:CURR:LIM:POS 3, (@1
For N6785A, N6786A
CAL:CURR:LIM:NEG -8, (@1)
Step 3.
Select the first current calibration point.
Front Panel:
SCPI Command:
The first calibration point is automatically
selected. “Enter P1 measured data” should be
displayed.
CAL:LEV P1
*OPC?
Step 4.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
SCPI Command:
Enter the calculated value into the data field
and then select Next.
CAL:DATA <data>
Step 5.
Select the second voltage calibration point. Wait 5 minutes for the
internal temperature to stabilize before continuing.
Front Panel:
SCPI Command:
The second calibration point is automatically
selected. “Wait 5 minutes, press Next” should
be displayed. Wait 5 minutes and select Next.
“Enter P2 measured data” should be displayed.
CAL:LEV P2
*OPC?
Step 6.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
SCPI Command:
Enter the calculated value into the data field
and then select Next. When calibration is
complete, select Close.
CAL:DATA <data>