3 Operations and Features
42 Keysight U8001A/U8002A User’s and Service Guide
To disable the OCP
1 Press on the “Over Current” button to disable the OCP. The OCP annunciator
will eventually turn off.
To clear the over current condition
1 OCP trip can only occur when the output is enabled. When the OCP condition
occurs:
The output is disabled and the OFF annunciator will be turned on.
If the keylock feature was enabled, it will be disabled. The LOCK annuciator
will be turned off.
The OCP annunciator will fast-blink continuously.
The display will show “TRIP”.
2 To clear the OCP trip conditon, press on the “Over Current” button again.
NOTE
If OCP trip persists, decrease the current limit settings to clear the tripping.