Service Guide 5
Keysight U8001A/U8002A User’s and Service Guide 77
Current Calibration
For current calibration, connect an appropriate shunt 0.01 Ω across the output 
terminals, and connect a digital voltmeter (DVM) across the shunt resistor. Same 
as voltage calibration, you will also be calibrating for low current, middle current 
and high current calibration point respectively. Make sure that the power supply is 
in constant current (CC) mode along the current calibration.
1 Current calibration mode
– Display will show “C” blinking.
– The CC annunciator will be turned on.
– Ensure that the power supply is in CC mode. Otherwise the CV annunciator 
will fast-blink.
2 Low current calibration point.
– Display will show “C” blinking.
– The M1 annunciator will be turned on.
– Use the knob to enter the computed value (DVM reading divided by shunt 
resistance).
– Press the “Memory” button to save the changes and move to the next 
calibration point. You will be able to perform this if the power supply is in 
CC mode.