Constant Current (CC) Verification
Current programming and readback accuracy
This test is to verify that the current programming and readback accuracy are 
within published specifications. The accuracy of the current monitoring resistor 
must be 0.1% or better.
Procedures:
1 Power-off the power supply and connect a 0.01 
Ω current shunt monitoring 
resistor (R
M
) across the output to be tested and a DMM across the current 
monitoring resistor (R
M
).
2 Power on the power supply.
3 When the power supply is in limit mode, program the output voltage to 
maximum programmable value and the current to 0 A.
4 Enable the output. 
5 Wait for a few seconds for the output of the power supply to settle. Make sure 
that the power supply is in CC mode.
6 Divide the voltage drop (DMM reading) across the current monitoring resistor 
(R
M
) by its resistance to convert to amps and record this value (I
O
). This value 
should be within the limit of 0 A ±20 mA.
7 When the power supply is in meter mode, record the current reading displayed 
on the front display of the power supply. This value should be within the limit 
of I
O
 ±20 mA.
8 Disable the output.
9 When the power supply is in limit mode, program the output current to full 
rated value, i.e. 3 A for U8001A and 5 A for U8002A.
10 Enable the output.
11 Wait for a few seconds for the output of the power supply to settle. Make sure 
that the power supply is in CC mode.