7.6 Setting up PC-KIT-NET (optional)
Tab. 7: Factory settings TCP/IP
IP address of unit 192.168.0.101
Subnet mask 255.255.255.0
IP address of standard gateway 192.168.0.200
If you want to commission more than one unit, the fol-
lowing applies:
Always connect one unit after the other to the network
and complete the network connection on the PC.
The IP address of the unit and the IP address of the
standard gateway must be configured in the same
address field:
IP address of unit (example) 180.160.15.1
IP address of standard gateway
(example)
180.160.15.2
For temperature monitoring via the network, the
KIRSCH-DATANET software (version 5.0 or higher)
must be installed (complete assembly and connection
manual is supplied with PC-KIT-NET).
Every unit needs its own IP address that is not yet in
use in your local network.
At the factory, the IP address is set to: 192.168.0.101
1. Insert unit key in key switch.
2. Set unit key to “P”.
3. Press and hold [Max.] and [Min.] simultaneously for four sec-
onds.
ð
The display shows .
4.
Press [Max.] or [Min.] to go to level and confirm with
Set.
Setting up the IP address on the
unit
Operation
Setting up PC-KIT-NET (optional)
23.05.2023Refrigerator/freezer MED series48