9 Maintenance
Interval Maintenance work
At least every 6 months
Check the condenser
Ä
Chapter 9.1 ‘Safety inspection’ on page 58.
9.1 Safety inspection
Perform safety inspection every two years!
The unit should be inspected by the owner in line with
DGUV regulation 3 (formerly BGV A3).
Subject the unit to a safety inspection at regular inter-
vals, but at least every two years.
The safety inspection performed by the owner contains the fol-
lowing individual inspections and their documentation:
1 - Visual inspection
2 - Function check
3 - Temperature check
4 - Temperature alarm test
5 - Inspection of the condenser
NOTICE!
If there is any doubt that the unit is in the correct condi-
tion, immediately stop using the unit. To prevent unin-
tended use, label the unit accordingly. Contact the
service department
Ä
‘Service contact’ on page 4.
1. Check the entire unit for completeness, correct set-up and
damage.
2. Check the following parts of the unit individually for damage:
n Door handle
n Interior
n Door seal
3. If there is damage and functionality is not ensured, decom-
mission the unit and contact the service department
Ä
‘Service contact’ on page 4.
Contents of the safety inspection
Visual inspection
Maintenance
Safety inspection
23.05.2023Refrigerator/freezer MED series58