KODAK RVG 6500 System_Regulatory User Guide (SM737)_Ed01B
13
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The KODAK RVG 6500 with or without IPS System is intended for use in the
electromagnetic environment specified below. The customer or the user of the KODAK
RVG 6500 with or without IPS System should assure that it is used in such an
environment.
Immunity Test
IEC 60601
Test Level
Compliance
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s)
to line(s)
Not applicable
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in
UT)
for 0,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 sec
<5% UT
(>95% dip in
UT)
for 0,5 cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
(30% dip in
UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the KODAK
RVG 6500 with or without IPS System
requires continued operation during
power mains interruptions, it is
recommended that the KODAK RVG
6500 with or without IPS System be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.