OPERATORS MANUAL
14 19JAN99 – 7C8770
Check that the correct Bulb and Safelight Filter are in the Safelight and at the
correct distance from the Feed Shelf and work surface.
Check that the Top Cover and Panels are tight on the Processor. Check that
there are no leaks in the Lighttight Gasket.
10 x 10 cm films - feed films diagonally if they fail to transport reliably.
Check the time delay. For all transport speeds, the buzzer should sound once
the trailing edge of the film has advanced 75 mm (3 in.) into the Processor.
Check that the tank solution levels are at the overflow Weirs.
Check for solution in the Replenishment Tanks. Fill if necessary.
NOTE: Mix developer replenisher in quantities not to exceed a 2-week
supply.
With a thermometer of known accuracy, check that the temperature of the
developer is correct. If necessary, have qualified service personnel make
adjustments.
NOTE: Check the incoming water temperature. It must be a minimum of 4 C
(7 F) lower than the desired developer temperature.
1. Transport Failure
2. Surface Artifacts
3. Abnormal Film Densities
4. Wet Films
5. Low Solution Levels
6. Overlapping of Films
123456