OPERATORS MANUAL
16 19JAN99 – 7C8770
Every Two Weeks
Figure 12 Biweekly Cleaning of the Racks and
Crossover Assemblies
Warning
Dangerous Voltage. Disconnect the main power before
doing any of the following cleanup procedures.
[1] Turn off the Main Power Switch.
[2] Disconnect the main power.
[3] Turn off the water supply.
[4] Remove the Top Cover.
Caution
When removing the Crossover Assemblies, be
careful to maintain squareness, and do not disturb
the Guide Shoes.
Use a synthetic sponge and warm water. Do not
use any abrasive material on the Racks,
Crossover Assemblies, or Squeegee Rollers.
Do not rinse the Detector Crossover Assembly or
the Dryer Assembly.
[5] Remove the Evaporation Covers and the
Crossover Assemblies.
(a) Rinse the Evaporation Covers.
(b) Rinse the developer/fixer and fixer/wash
Crossover Assemblies in warm water.
[6] Rotate and clean the Rollers with a soft brush and
warm water.
[7] Dry the Rollers with a damp cloth or synthetic
sponge.
[8] Wipe off all chemical deposits above the solution
level in the processing section.
[9] Remove the Dryer Assembly.
[10] Clean off debris or buildup from the dryer area and
Dryer Assembly with a warm, damp sponge or
cloth.
H172_1001CC
H172_1001CCB
SHELF
FEED
GUIDE
SHOE
DETECTOR
CROSSOVER
ASSEMBLY
CROSSOVER
ASSEMBLIES
DRYER
ASSEMBLY