Replenishment Flow Rates
7C8770 – 19JAN99 11
Section 4: Replenishment Flow Rates
Figure 9
Warning
Wear protective eyewear. The replenishment solutions
are pumped quickly and might splash.
[1] Remove the Top Cover.
[2] Lift the Upper Roller of the Detector Crossover
Assembly.
[3] Check that the replenisher solutions flow freely
through the Tubing located along the drive side of
the Racks.
[4] Disconnect the Developer Tubing by pressing the
Metal Latch.
Note
The Developer Tubing may have a red Wire Tie on it for
easy identification.
[5] Pull the Developer Tubing slightly and rotate it
over the edge of the frame and into a Graduated
Cylinder.
[6] Lift the Upper Roller of the Detector Crossover
Assembly for the correct time:
[7] When the Replenishment Pump stops, check that the amount of solution in the Graduated Cylinder is correct
for the film size and use condition. See the table on Page 12. If the amount is not correct, contact qualified
service personnel.
[8] Connect the Developer Tubing again by pushing it into the Metal Latch until the Developer Tubing locks in place.
[9] Check the fixer replenishment flow rate by doing Steps 3 - 8 with the Fixer Tubing.
[10] Install the Top Cover.
Figure 10 Detector Crossover Assembly
H172_1005AC
H172_1005ACA
DISCONNECT
QUICK
GRADUATED
FIXER TUBING
CYLINDER
METAL LATCH
DEVELOPER TUBING
Standard Cycle: 34 seconds
Rapid Cycle: 25 seconds
H172_1006BC
H172_1006BCA
SWITCHES
2 DETECTOR
ASSEMBLY
DETECTOR CROSSOVER
UPPER ROLLER