Cleanup and Periodic Maintenance
7C8770 – 19JAN99 15
Section 6: Cleanup and Periodic Maintenance
Daily Procedure
The best time to do daily cleanup is at shutdown.
Figure 11 Daily Cleaning of the Racks and
Crossover Assemblies
Warning
Dangerous Voltage. Disconnect the main power
before doing any of the following cleanup
procedures.
Wear rubber gloves, safety glasses and protective
clothing when doing any maintenance procedure.
Report any change in the operating condition of the
Processor to Service Personnel.
[1] Turn off the Main Power Switch.
[2] Disconnect the main power.
[3] Turn off the water supply.
[4] Remove the Top Cover.
Caution
When removing the Crossover Assemblies, be
careful to maintain squareness, and do not disturb
the Guide Shoes.
Use a damp cloth or a synthetic sponge and warm
water. Do not use any abrasive material on the
Racks, Crossover Assemblies, or Squeegee
Rollers.
[5] Remove the Evaporation Covers and the
Crossover Assemblies.
(a) Rinse the Evaporation Covers.
(b) Rinse the developer/fixer and fixer/wash
Crossover Assemblies in warm water.
[6] Rotate and clean the Rollers with a soft brush and
warm water.
[7] Dry the Rollers with a cloth or synthetic sponge or
allow to air dry.
[8] Wipe off all chemical deposits above the solution level in the processing section.
[9] Clean off debris or buildup from the dryer area with a warm, damp sponge or cloth.
[10] Install the Crossover Assemblies and the Evaporation Covers.
Important
Keep the Top Cover open overnight for venting purposes.
[11] Install the Top Cover leaving a 5 cm (2 in.) gap.
H172_1001CC
H172_1001CCB
SHELF
FEED
GUIDE
SHOE
DETECTOR
CROSSOVER
ASSEMBLY
CROSSOVER
ASSEMBLIES
DRYER
ASSEMBLY