Cleanup and Periodic Maintenance
7C8770 – 19JAN99 15
Section 6: Cleanup and Periodic Maintenance
Daily Procedure
The best time to do daily cleanup is at shutdown.
Figure 11 Daily Cleaning of the Racks and
Crossover Assemblies
Warning
• Dangerous Voltage. Disconnect the main power
before doing any of the following cleanup
procedures.
• Wear rubber gloves, safety glasses and protective
clothing when doing any maintenance procedure.
• Report any change in the operating condition of the
Processor to Service Personnel.
[1] Turn off the Main Power Switch.
[2] Disconnect the main power.
[3] Turn off the water supply.
[4] Remove the Top Cover.
Caution
• When removing the Crossover Assemblies, be
careful to maintain squareness, and do not disturb
the Guide Shoes.
• Use a damp cloth or a synthetic sponge and warm
water. Do not use any abrasive material on the
Racks, Crossover Assemblies, or Squeegee
Rollers.
[5] Remove the Evaporation Covers and the
Crossover Assemblies.
(a) Rinse the Evaporation Covers.
(b) Rinse the developer/fixer and fixer/wash
Crossover Assemblies in warm water.
[6] Rotate and clean the Rollers with a soft brush and
warm water.
[7] Dry the Rollers with a cloth or synthetic sponge or
allow to air dry.
[8] Wipe off all chemical deposits above the solution level in the processing section.
[9] Clean off debris or buildup from the dryer area with a warm, damp sponge or cloth.
[10] Install the Crossover Assemblies and the Evaporation Covers.
Important
Keep the Top Cover open overnight for venting purposes.
[11] Install the Top Cover leaving a 5 cm (2 in.) gap.
H172_1001CC
H172_1001CCB
SHELF
FEED
GUIDE
SHOE
DETECTOR
CROSSOVER
ASSEMBLY
CROSSOVER
ASSEMBLIES
DRYER
ASSEMBLY