Overview
7C8770 – 19JAN99 3
Section 1: Overview
The
Kodak X-Omat
2000 and 2000A Processors are designed to process medical x-ray sheet or roll film that can be
processed in
Kodak RP X-Omat
Developer and Replenisher and in
Kodak RP X-Omat
Fixer and Replenisher.
The self-threading roller transport system has:
• a film-length detection system
• a replenishment system
• a developer solution filter
• a recirculation system for the developer and fixer solutions
• an automatic standby feature
The Processor can operate at 2 different transport speed cycles:
• Standard Cycle has a transport speed of 76.2 2.0 cm (30 0.6 in.) per minute or 157 5 seconds drop time
(dry to dry) for a 43 cm (17 in.) sheet of film. The developer temperature is 33.3 0.3°C (92.0 0.5°F).
• Rapid Cycle has a transport speed of 101.6 1.5 cm (40.1 0.8 in.) per minute or 118 4 seconds drop time
(dry to dry) for a 43 cm (17 in.) sheet of film. The developer temperature is 34.4 0.3°C (94.0 0.5°F).
• Both Cycles use standard
Kodak RP X-Omat
Fixer and Replenisher.
Note
The transport speed can only be changed by qualified personnel.
A film feed signal sounds 3 seconds after the film being processed has cleared the Rollers of the Detector Crossover
Assembly. This indicates that the Processor is ready to accept another piece of film.