Cleanup and Periodic Maintenance
7C8770 – 19JAN99 17
Figure 13 Silver Recovery Unit
Caution
Fixer will contaminate the developer. When
removing the Rack Assemblies, use the Splash
Guard between the developer and fixer Tanks.
Use the Drip Tray when removing or installing the
Racks. See Figure 6 on Page 8.
Do not rinse the Detector Crossover Assembly or
the Dryer Assembly.
[11] Remove the Racks.
[12] Rinse the Racks and wipe them with a cloth.
[13] Check:
(a) Chains move freely and are adjusted
correctly
(b) Rollers rotate correctly
(c) Space between the Turnaround Side Plates
and the Rack Side Plates is equal and that
the Side Plates are parallel on both ends
Figure 14 Rack and Turnaround Assembly
[14] Check the operation of the Silver Recovery Unit, if
installed.
[15] Check that no solution is flowing through the
Bypass.
Note
If solution is present in the Bypass, the Silver Recovery
Unit is not operating.
[16] Use
Kodak
Silver Estimating Test Papers to check
the silver content of the solution flowing into the
drain.
[17] If the test shows more than 1 gram/liter of silver,
install a new cartridge in the Silver Recovery Unit.
H172_1007AC
H172_1007ACA
SILVER RECOVERY UNIT
BYPASS
H172_1008AC
CHAIN
ROLLER
STUD
2 RACK
SIDE PATES
2 TURNAROUND
SIDE PLATES
H172_1008ACB