Konica Minolta CS-7 - AEC Performance Checks (Stability and Reproducibility

Konica Minolta CS-7
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
169
4.2 MP-Tables
Chapter 4
3.13 AEC Performance Checks
Table MP-10 (CNR)
Date: Inspectors Name:
Mammography X-ray unit S/N:
Stability and Reproducibility
28kVp, 4 cm acrylic
Exp Count
mAs mR
Exp #1
Exp #2
Exp #3
Exp #4
Mean
Standard Deviation (SD) Pass/Fail
Coeff of variation (%)
*1
Criteria
≤5% ≤5%
P Pass
*1Coeff of variation(%)=(SD/mean)×100 F Fail
Thickness Tracking
Thickness
(cm)
kVp Target Filter
mAs
CNR
Measured
CNR
(relative to
4cm PMMA)
Acceptable
level
Pass/Fail
2 >100%
4 100%
6 >75%
P Pass
F Fail
Comments: