Variable area flowmeter
Equipment category II 2 G and II 2 D, EPL Gb and Db
in protection type intrinsic safety "i"
DK37/M8 - H250/M8
DK37/M8 - H250/M8DK37/M8 - H250/M8
DK37/M8 - H250/M8
Supplementary Instructions
Supplementary InstructionsSupplementary Instructions
Supplementary Instructions
© KROHNE 05/2018 - 4005351602 - AD DK37/H250/M8 II2GD R02 en
Question and Answer IconNeed help?

Do you have a question about the KROHNE H250 Series and is the answer not in the manual?

KROHNE H250 Series Specifications

General IconGeneral
Accuracy±1.6%
Process PressureUp to 40 bar
MaterialsStainless steel, PTFE, Hastelloy
Output4...20 mA, HART
DisplayLCD
Process ConnectionFlange
ApprovalsATEX, IECEx, FM, CSA
Measurement PrincipleVariable area (VA) flow meter

Summary

Safety Instructions

General Notes

Applies to explosion-protected versions, completing standard documentation for non-explosion protected types.

EU Conformity

Declares conformity with ATEX directive and harmonized standards for hazardous areas.

IECEx Scheme Approval

Details testing and certification according to the IECEx scheme for hazardous areas.

Safety Instructions

Warns about explosion risk if instructions are not followed and requires trained personnel.

Device Description

Device Description

Explains the function of variable area flowmeters for flammable and non-flammable gases and liquids.

Description Code

Explains the safety description code for type series H250, listing its components.

DK37 Description Code

Explains the safety description code for type series DK37, listing its components.

Marking

Details the markings on the device nameplate and housing cover for identification and compliance.

Flammable Products

Discusses atmospheric and operating conditions relevant to flammable products and ATEX directives.

Equipment Category

Explains equipment categories II 2 G and II 2 D for use in Zone 1 and Zone 21 respectively.

Types of Protection

Details intrinsic safety protection ("i") as per EN 60079-11 for gas and dust environments.

Ambient/Product Temperature

Explains how product and ambient temperatures influence the device's temperature class and operational limits.

Surface Temperature (II 2 D)

Notes maximum surface temperature for dust environments (II 2 D), valid without dust coating.

Electrical Data

Lists maximum values (U, I, P, C, L) for intrinsically safe circuit connections.

Installation

Mounting

Covers mounting requirements, installation standards (EN 60079-14), and accessibility for explosion protection.

Special Conditions

Discusses electrostatics, titanium components for oxygen applications, and thermal/electrical data considerations.

Electrical Connections

General Notes

Details intrinsically safe and galvanically isolated circuits, cable selection, and laying requirements.

Power Supply

States that power for built-in equipment is supplied via the signal circuits.

Inputs/Outputs

Refers to product documentation for terminal assignment and requires connection to certified intrinsically safe circuits.

Grounding and Bonding

Explains establishing ground connections via terminal ① and insulating unused cores for safety.

Operation

Start-up

Outlines conditions for permitted start-up, including correct installation and system checks.

Operation

Covers operating within specified temperature, pressure, and electrical limits.

Electrostatic Charge

Warns about electrostatic hazards with painted versions and plastic housings in specific environments.

Service

Maintenance

Details requirements for safety-relevant maintenance by qualified personnel and recommended checks.

Dismantling

Covers replacing the display and exchanging the entire device, including safety precautions.