EasyManua.ls Logo

Kubler 571T - Display Menu

Kubler 571T
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R60723.0002–Index6c EN‐page46/62
5.15 DisplayMenu
Parameterchangesbecomeactiveonlyafterclosingthemenuselection.
SOURCEDUALTOP
(Referencesourcefortwo‐linedisplay,firstline)
0
MEASUREM.
RESULT
Referencesourceisthemeasurementresultoftheselected
operatingmode.
1 COUNTERA
ReferencesourceisthecounterreadingofchannelA.
(onlyvisibleinmodeCOUNTERCOUNTMODE:A+BorA‐B)
2 COUNTERB
ReferencesourceisthecounterreadingofchannelB.
(onlyvisibleinmodeCOUNTERCOUNTMODE:A+BorA‐B)
3 BATCHCOUNTER
Referencesourceisthebatchcounter.
(onlyvisibleinmodeCOUNTERBATCH/TOTALMODE:
INCREMENTBATCH,DECREMENTBATCHorUSEINPUTSONLY)
4 TOTALCOUNTER
Referencesourceistotalcounter.
(onlyvisibleinmodeCOUNTERBATCH/TOTALMODE:TOTAL
COUNTER)
5 TOTALTIMER
Referencesourceisthetotaltimer.
(onlyvisibleinmodeTIMERTOTALTIMEMODE:YES)
SOURCEDUALDOWN
(Referencesourcefortwo‐linedisplay,secondline)
0
MEASUREM.
RESULT
Referencesourceisthemeasurementresultoftheselected
operatingmode.
1 COUNTERA
ReferencesourceisthecounterreadingofchannelA.
(onlyvisibleinmodeCOUNTERCOUNTMODE:A+BorA‐B)
2 COUNTERB
ReferencesourceisthecounterreadingofchannelB.
(onlyvisibleinmodeCOUNTERCOUNTMODE:A+BorA‐B)
3 BATCHCOUNTER
Referencesourceisthebatchcounter.
(onlyvisibleinmodeCOUNTERBATCH/TOTALMODE:
INCREMENTBATCH,DECREMENTBATCHorUSEINPUTSONLY)
4 TOTALCOUNTER
Referencesourceistotalcounter.
(onlyvisibleinmodeCOUNTERBATCH/TOTALMODE:TOTAL
COUNTER)
5 TOTALTIMER
Referencesourceisthetotaltimer.
(onlyvisibleinmodeTIMERTOTALTIMEMODE:YES)
COLOR
Thisparameterdefinesthedisplaycolor.
Event‐dependingchangeofthedisplaycolorbyaswitchingconditionispossible(seePRESELECTION
1…4MENU)
Event‐dependingchangesare
onlyavailablefordeviceswithoptionAnalog‐,Control‐,Relay‐Output.
0 RED Reddisplay
1 GREEN Greendisplay
2 YELLOW Yellowdisplay