Under normal operation, the CentriVap requires little
maintenance. The following maintenance schedule is
recommended:
As needed:
1. Clean up all spills; remove liquids from the
chamber. Clean or decontaminate all surfaces using
agents suitable for the substance spilled.
2. Clean lid and gasket using soft cloth, sponge or
chamois and a mild, non-abrasive soap or detergent.
3. Check oil level of the vacuum pump, if applicable.
It should be between MIN and MAX. If the oil
level is less than an inch (25.4 mm) above MIN, add
oil to proper level.
4. If oil shows cloudiness, particles or discoloration,
drain the pump and replace with fresh oil.
5. Utilization of acids requires immediate cleaning and
neutralization after a run or physical damage to the
collection chamber will result.
6. Check the Cold Trap for condensed or frozen
solvents and dispose of appropriately. Completely
empty the trap before the next run. If solvents are
frozen in the glass trap, run it under cold water
immediately after operating.