Refer to the following if your CentriVap fails to operate
properly. If the suggested corrective actions do not
solve your problem, contact Labconco for additional
assistance. The following failure codes may appear on
the display when problems are sensed by the internal
self check routine.
DISPLAY
ERROR CODE
CAUSE CORRECTIVE ACTION
HEAT SENSOR
Sensor failure Replace sensor assembly.
Connection failure Repair connection.
CLOSE LID
Lid open Close lid.
LATCH FAIL
Solenoid failure Check component.
Switch or sensor
failure
Check connections.
MEM FAIL P1
Memory failure Push program button #1
New memory IC chip Push program button #1.
Bad memory IC chip Call Labconco – Replace IC chip or control PCB.
MOTOR ERROR
Defective motor Replace motor.
Hall effect sensor
failure
Replace sensor.
Wire failure Replace wire.