Window motor board
8.5 Replacement of laminator
The laminator is placed at the top of the working chamber just below the main downflow filter. To
remove the laminator, the two fixing screws, shown on the picture below, must be removed.
Pull on the two flaps located at the back edge of the laminator and pull downwards. Be careful not
to damage the main downflow filter located just above the laminator.
Install the replacement laminator and mount the fixing screws again.
8.6 Replacements and repairs
All replacement and repairs – other than the two described above – must be performed by a trained
service technician. Any retro-fitting of optional equipment
8.7 Disposal of cabinet
The entire cabinet except for the HEPA filter can be discarded after having been thoroughly cleaned
and decontaminated. The HEPA filter must be discarded in accordance with the applicable national
and local regulations for chemical and biological waste.