9. To repeat the test, press
ENTER
to scan the sample again. The
last blank scanned is used by
the colorimeter for repeated
scans. A different blank can
be used by pressing or
to scroll to Scan Blank
and then scanning another
blank. Scroll with
or
and make another
selection with
ENTER
.The %T or
Absorbance of the last test can
be viewed by choosing
%T/Abs. Press
EXIT
to
escape to previous menus.
NOTE: The menus loop in
this screen so either or
will lead to the menu
selection needed.
NOTE: The calibrate function
does not work in the
Absorbance mode.
102 Absorbance 525
0.425
Scan Blank
Scan Sample
12:00:00 001/500
Next Test
Previous Test
%T/Abs
Calibrate
SMART3 Colorimeter 11.10 35
Testing