5 Maintenance
5.1.7 Premature Cold Start Termination
Occasionally, it might be necessary to terminate the COLD START prematurely, e.g., after a
short-term power failure, but not before a minimum waiting period of 120 seconds has expired.
NOTICE
A premature cold start termination leads directly into measurement operation without an addi-
tional prompt. If the probe does not reach its optimum operating point, falsified measured val-
ues and, possibly, faults and warnings are the result.
In the main menu, select for the calibra-
tion of the probe.
Interrupt the cold start prematurely.
By selecting with the cursor key and
ENTER the cold start is terminated pre-
maturely.
The system returns to the main menu auto-
matically.