EasyManua.ls Logo

Larson Davis Spartan 821 - 3.5 Calibration Settings

Larson Davis Spartan 821
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spartan 821/721 Manual Calibration Settings 28
3.5 Calibration Settings
FIGURE 3-4 Calibration Card
Level
Set the Calibrator output level.
System Sensitivity
Set the system sensitivity either in dB re 1V/Pa or
mV/Pa.
Correction
Select the correction filter. See A.1.4
Microphone Model 377B02 Filter Response
and Corrections
Auto Calibration
Enable the meter auto calibration feature, which
will detect when a calibrator tone is injected
around the calibrator level while reset or
stopped.
Microphone settings
Select the microphone attached to the system
and enter it’s serial number to be stored with
measurement data. Custom microphone
information can be entered using the Other”
model option
Preamp Serial Number
Enter the Preamp serial number. Supported
Larson Davis preamp types are detected
automatically

Related product manuals