Integral T process thermostats
21/01/2019 / YAWE 0026 Preparations 27
Risk to the environment due to oil contamination of the cooling water circuit
With a leaky condenser there is the danger that refrigerating machine oil from the refrigerant circuit of the
cooling thermostat can pass into the cooling water.
Follow all the legal requirements and the regulations of the water supply utility which apply at the point of
use.
Water pollution due to leakage
To avoid pollution due to a leak in the cooling water system it is recommended that a leakage-water
detector with a water cut-off is installed.
Servicing intervals
Follow the information for cleaning and decalcifying the cooling water circuit ( 8.6).
inner diameter
Ø mm x insulation
thickness
maximum
working
pressure in bar
Tubing with
reinforcement
EPDM
for all heat transfer
liquids except Ultra 350
and mineral oils;
all units with ½“ nipple
Tubing with
reinforcement
EPDM
for all heat transfer
liquids except Ultra 350
and mineral oils;
all units with ¾“-nipple
Tubing with
reinforcement
EPDM
for all heat transfer
liquids except Ultra 350
and mineral oils;
all units with 1“-nipple
inner diameter
Ø mm x wall
thickness
EPDM tubing is not suitable for Ultra 350 and not suitable for mineral oils.
Silicone oil causes pronounced swelling of silicone rubber. Never use silicone oil with
silicone tubing!
Protect tubing with hose clips against slipping off!