20 / 52 Alpha 7/2022
When operating as bath thermostat without external consumer the pump outflow connec-
tion has to be linked to the return. At bath temperature above 70 °C the label HOT sur-
face supplied must be affixed on the bath at clearly visible position!
The unit can be operated safely up to an ambient temperature of 40 °C.
Depending on the loading of the refrigeration system, a temporary shut-off can occur, especially in
case of an ambient temperature of over 35 °C.
Additionally a higher ambient temperature results in less refrigerating capacity.
When starting up the refrigeration system after a longer time, it can take up to 30 minutes, de-
pending on the ambient temperature and the unit type, until the nominal refrigerating capacity is
reached.